Continuous Blender Troubleshooting
For troubleshooting purposes,
the Process Control blender should be viewed as two separate
systems:
-
the Weigh System which precisely
weighs the material and transmits the readings to the operator
station
-
the Feeder Drive System which
receives speed change commands from the operator station and
controls the feeders accordingly.
Some hints about checking each
of these systems as well as an important point about grounding
flex hose can be found below.
The Weigh System
In order to produce accurate
blends, the weigh system must be able to accurately measure the
amount of material being removed from the hopper (the weight
loss). It is best to check the weigh system while the blender
is paused. Each hopper weight can be viewed on the display using
MONITOR. No weight fluctuations should be seen during a paused
condition. Any fluctuation indicates either:
-
Hopper mechanical interference.
This is the most common cause of inaccurate weight readings.
Inspect each hopper very closely to see that nothing is touching
the hopper. Common points of mechanical interference are:
-
the end of the auger
tube touching the sample chute or cascade chamber;
-
the loadcell overload protection
screws touching the loadcell;
-
the hopper drain tube touching
the blender loadcell platform;
-
flexible hoses or material lines
touching the hoppers. These points are normally in close proximity
to the weigh hopper, but should not touch unless something has
shifted during shipment or use.
-
Nonlinearity of the weigh electronics
or loadcell. Linearity is described as the weigh systems ability
to measure weight accurately over the entire range of the loadcell.
To perform this test, view the hopper weight on the display using
MONITOR. The hopper may have material in it. Hang a calibration
test weight on the hopper and verify the hopper weight rises
by exactly that amount. A hopper which reads 12.25 pounds of
material prior to hanging a 10 pound test weight should now indicate
22.25 pounds. Any deviation more than 0.01 pounds from the target
value indicates the need for hopper calibration or possibly a
defective loadcell. Check with PCC Service for a quick way to
test a load cell.
The Feeder Drive System
The feeder drive system for each
hopper should be tested for stability and linearity. This is
best accomplished with the weigh hopper paused and emptied of
material. The user must first enable the manual backup mode on
the blender. This will allow for manual speed setting of the
feeder motor being tested, which can be verified using a hand
held tachometer positioned directly onto the drive end of the
auger shaft.
The user should first set the
manual speed setting to 100% and verify the auger RPM indicates
the motor gearbox maximum RPM. The user would then change the
manual speed setting to 50%, 25% and finally 10% motor speeds,
all the while measuring and recording the auger shaft RPM. The
measured values should be directly proportional to the percentage
of the maximum motor gearbox RPM. Additionally, the measured
RPM values should be stable during the test. An RPM value which
fluctuates by plus or minus 2 RPM is not stable and could cause
poor metering performance during gravimetric operation.
Ground Material Flex Hose (see
diagram)
One common cause of blender performance
problems is improper grounding. The static charges produced by
the material moving across the inside of the flex hose can be
enormous. This is especially noticeable when humidity is low
allowing high-voltage arcs to occur. Grounding of material lines
should be taken very seriously. Failure to properly ground material
flex hoses is a hazard to operator personnel and electronic equipment.
Material hoses must be grounded to a good earth ground located
AWAY from the blender and receivers. This includes material lines
positioned vertically between a surge bin and knifegate assembly.
Remote mezzanine mounted blenders cascading material over a vertical
distance through flex hose must utilize grounded hose as well.
Material flex hoses supplied by PCC include a static discharge
wire enclosed in the flex hose length. This wire must be stripped
back and grounded to the fixed material line or some supporting
structure which is connected permanently to a good earth ground.
At no time should this wire to be attached to anywhere on the
receiver or blender as this simply discharges the static directly
into the receiver and blender frame causing likely damage to
the electronics. Installations utilizing material hose without
a static ground wire can be wrapped with a bare braided wire
like a barber pole or candy cane over the entire length of the
material hose. Hose clamps are suitable for holding such a wire
in place and again the wire must be terminated to a good earth
ground away from the blender and receivers.
Additional Troubleshooting Tips:
Additional troubleshooting tips are available in each of our equipment manuals. Visit Manuals and Software to view these items.